Key result
Outlet strut fractures occurred in 0.7% of 86,000 implanted Björk-Shiley convexoconcave valves over 25 years, with current rates in 60-degree valves at <0.1% per year.
Why the study?
What are the long-term risks of outlet strut fracture and the indications for prophylactic reoperation in patients with Björk-Shiley convexoconcave heart valves?
What are the long-term risks of outlet strut fracture and the indications for prophylactic reoperation in patients with Björk-Shiley convexoconcave heart valves?
Twenty-five years post-implantation, Björk-Shiley convexoconcave valves continue to carry a small but persistent risk of outlet strut fracture, necessitating individualized risk assessment for prophylactic reoperation.
BACKGROUND: The first Björk-Shiley convexoconcave (BSCC) prosthetic heart valves were implanted in 1978. The 25th anniversary provided a stimulus to summarize the research data relevant to BSCC valve fracture, patient management, and current clinical options. METHODS AND RESULTS: Published and unpublished data on the risks of BSCC valve fracture and replacement were compiled, and strategies for identifying candidates for prophylactic valve reoperation were summarized. By December 2003, outlet strut fractures (OSFs), often with fatal outcomes, had been reported in 633 BSCC valves (0.7% of 86,000 valves implanted). Fractures still continue to occur, but average rates of OSFs in 60 degrees valves are now <0.1% per year. OSF risk varies markedly by valve characteristics, especially valve angle and size, with weaker effects associated with other manufacturing variables. OSF risks are mildly lower among women than men but decline sharply with advancing age. The risks of valve replacement typically greatly exceed those of OSF. By comparing individualized estimated risks of OSF versus valve replacement, guidelines have been developed to identify the small percentage of BSCC patients (mostly younger men) who would be expected to have a gain in life expectancy should reoperative surgery be performed. CONCLUSIONS: Twenty-five years after the initial BSCC valve implants, fractures continue to occur. Continued monitoring of BSCC patients is needed to track and quantify risks and enable periodic updating of guidelines for patients and their physicians.
No takes yet. Share an insight, caveat, or question.
Blot et al. (2005) conducted a review in Björk-Shiley convexoconcave (BSCC) prosthetic heart valves (n=86,000). Björk-Shiley convexoconcave (BSCC) prosthetic heart valves was evaluated on Outlet strut fractures (OSFs). Outlet strut fractures occurred in 0.7% of 86,000 implanted Björk-Shiley convexoconcave valves over 25 years, with current rates in 60-degree valves at <0.1% per year.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: