Key result
Large Björk-Shiley convexo-concave heart valves (≥29 mm) were associated with a 6.7-fold increased risk of crack or single-leg fracture compared with smaller valves (OR 6.7; 95% CI 1.8-24.6; P=0.04).
Observational (n=62)
A significant proportion of intact explanted Björk-Shiley valves have serious prefracture defects, with larger valves at higher risk, highlighting the difficulty of predicting fracture.
Answers are grounded in this paper and cite their sources.
May support enhanced surveillance for patients with large Björk-Shiley valves; leaves open prospective validation of fracture risk models.
OBJECTIVE: Non-destructive assessment of 62 Björk-Shiley convexo-concave heart valves in view of the risk of fracture and possible detection of defects by means of X-ray or acoustics. METHODS: Scanning electron microscopy and stereomicroscopy of an unselected sample of valves, representing approximately 70% of the valves explanted between 1991 and 1996. Mean duration of implantation was 10.7 +/- 1.9 years. RESULTS: Six of the 62 valves had a fracture with disk escape. Of the remaining 56 valves, 11 (19.6%) had a single-leg fracture and seven (12.5%) showed a crack. Valves, 24 (42.8%), revealed no significant defect. After 4600 days of implantation, large valves (i.e. > or = 29 mm) showed a 6.7 time increase in the risk of crack or single-leg fracture compared with smaller valves (OR 6.7, 95% CI 1.8-24.6, P = 0.04). In a small in-vitro experiment with single-leg fracture, rapid smearing occurred, resulting in strut leg separation upon some movement. CONCLUSIONS: After approximately 400 x 10(6) cycles, 42.5% of the retrieved intact valves was without significant defect, while 32% had a serious prefracture defect. Time evolution from defect to fracture remains unpredictable. Also, detection of prefracture defects, including single-leg fracture, remains difficult, because a "well-detectable' fracture with dislocation is just a brief condition. However, technical knowledge can enhance the epidemiological and manufacturing information used for clinical decision making.
No takes yet. Share an insight, caveat, or question.
Bas A.J.M. de Mol (1997) conducted an observational in Explanted Björk-Shiley convexo-concave heart valves (n=62). Large valves (≥ 29 mm) vs. Smaller valves was evaluated on Crack or single-leg fracture (OR 6.7, 95% CI 1.8-24.6, p=0.04). Large Björk-Shiley convexo-concave heart valves (≥29 mm) were associated with a 6.7-fold increased risk of crack or single-leg fracture compared with smaller valves (OR 6.7; 95% CI 1.8-24.6; P=0.04).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: