Key result
Acoustic detection of Björk-Shiley valve defects fails due to thoracic sound attenuation and variable excitation.
Why the study?
Fractures in Björk‐Shiley convexoconcave heart valves raised questions about the feasibility of early diagnosis of technical defects by acoustic assessment.
Acoustic assessment should not guide Björk-Shiley valve monitoring; leaves open validated noninvasive alternatives.
Fractures in Björk‐Shiley convexoconcave (BScc) heart valves have raised questions about the feasibility of early diagnosis of technical defects by means of acoustic assessment. Three laboratory tests were conducted. To establish acoustic fingerprints, 66 valves with a defect, such as single‐leg fracture (SLF) or single‐leg separation (SLS), or without a defect were connected with a contact sensor and excited by dropping a small metal ball onto the outlet strut. In the second test, we simulated the valve sound propagation within the thorax. In the third test, intact, SLF, and SLS valves were placed in a mock heart immersed in a large water tank. We observed a resonance frequency corresponding with valve size and presence of defects. The second test showed that both the chest wall and the lungs created numerous reflections. This led to a substantial overlap of the original pulse frequencies and the frequencies measured. The third test confirmed that submersion of the chest in water can significantly reduce chest wall reflections. Reliable noninvasive assessment of BScc valve clicks for the presence of defects of the outlet strut is hampered by complex sound propagation within the thorax and variability of valve excitation. Acoustic fingerprints to diagnose mechanical defects should be integrated in valve design.
No takes yet. Share an insight, caveat, or question.
Mol et al. (2001) studied Björk-Shiley convexoconcave heart valve defects (n=66). External acoustic assessment was evaluated on Detection of resonance frequency corresponding with valve size and presence of defects. Reliable noninvasive acoustic assessment of Björk-Shiley convexoconcave valve defects was hampered by complex sound propagation within the thorax and variability of valve excitation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: