Key result
Large Björk-Shiley convexo-concave valves linked to ~218% higher 10-year strut fracture risk versus small valves.
Why the study?
The risk and outcomes of outlet strut fractures in Björk-Shiley 70 degrees convexo-concave prosthetic heart valves implanted internationally were not fully characterized.
Does valve size affect the risk of strut fracture in patients with Björk-Shiley 70 degrees convexo-concave prosthetic heart valves?
Population
831 patients implanted with Björk-Shiley 70 degrees convexo-concave valves at five institutions in Sweden, Germany, The Netherlands, and Canada
Comparison
Large (29-33 mm) valves vs small (21-27 mm) valves
Design
International multi-institutional cohort study
Follow-up
Up to 10.1 years (median 4.6 years)
Authors
Loading...
Large valves were associated with higher strut fracture risk; extends prior data but leaves optimal surveillance or reoperation thresholds open.
Cohort (n=831)
Yes
Does valve size affect the risk of strut fracture in patients with Björk-Shiley 70 degrees convexo-concave prosthetic heart valves?
Absolute Event Rate: 10.5% vs 3.3%
p-value: p=<0.001
Large Björk-Shiley 70 degrees convexo-concave valves have a significantly higher risk of strut fracture compared to smaller valves, prompting recommendations for elective reoperation in high-risk patients.
Ericsson et al. (1992) conducted a cohort in Implanted Björk-Shiley 70 degrees convexo-concave prosthetic heart valves (n=831). Large (29-33 mm) Björk-Shiley 70 degrees convexo-concave prosthetic heart valves vs. Small (21-27 mm) valves was evaluated on 10-year actuarial strut fracture rate (p=<0.001). Large (29-33 mm) Björk-Shiley 70 degrees convexo-concave valves had a significantly higher 10-year actuarial strut fracture rate compared to small (21-27 mm) valves (10.5% vs 3.3%, P<0.001).
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: