This is the authors' abstract. We don't add key points for this paper.
To achieve optimal image quality and highest spatial resolution for inner ear imaging with a 3D gradient echo sequence, it is necessary to minimize susceptibility dephasing effects by using very short TE. Fractional RF pulses and echoes can yield short TE for moderate spatial resolution; however, for voxel size of less than 1 mm, TE is limited by the phase encode gradients. We present a method to obtain very short effective TE by using short triangular shaped phase encode gradients to sample the central portions of k-space and progressively longer trapezoidal gradients for the outer portions of k-space. A 3D pulse sequence employing the modified phase encoding scheme for both in-plane and slice phase encoding directions was implemented and tested on phantoms and in vivo. The effective TE equals the minimal TE used for the central k-space portions. Submillimeter resolution (0.35 x 0.35 x 0.7 mm3) images of the inner ear were obtained with effective TE of 3.2 ms and were compared with standard 3D images with TE of 8 ms. With this pronounced TE reduction, the susceptibility artifacts at air/fluid interfaces are significantly reduced.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Schmalbrock et al. (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: