Answers are grounded in this paper and cite their sources.
A precise, rapid, automated, rate nephelometric immunoassay for apolipoprotein AI (APA) and apolipoprotein B (APB) is described. Both analytes are determined by a "one-pot" procedure which uses consecutive addition of the corresponding antibodies. Poly(oxyethylene) type nonionic surfactants are used to selectively inhibit the APB reaction after the maximum reaction rate has been reached and to enhance the APA immunoreactivity. The assay range for APA is 0.3-3.8 g/L and for APB 0.2-3.0 g/L. The new assay was compared to commercial rate nephelometric methodology. Both methods were shown to be mutually unbiased for APA and APB determinations in human serum.
No takes yet. Share an insight, caveat, or question.
Fotis Fotiou (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: