Key result
Transabdominal ECG provided lower average signal loss than Doppler (5.3% vs 15.5%) but artificially increased fetal heart rate variability during active maternal positions (23.6 vs 13.5 bpm, P<0.001).
Why the study?
Continuous wireless TAfECG has enabled fetal heart rate monitoring during maternal mobilization, but its signal quality and accuracy across static and active labor positions needed evaluation against Doppler and fetal scalp electrodes.
Does transabdominal fetal ECG provide accurate fetal heart rate monitoring compared to Doppler and fetal scalp electrode during maternal movement in labor?
Observational (n=76)
Does transabdominal fetal ECG provide accurate fetal heart rate monitoring compared to Doppler and fetal scalp electrode during maternal movement in labor?
While TAfECG provides reliable fetal heart rate monitoring in static positions, it artificially increases FHR variability during maternal movement, which may lead to false reassurance.
May inflate FHR variability during maternal movement, risking misinterpretation; leaves open accuracy versus scalp electrode in active labor.
INTRODUCTION: Transabdominal electrocardiographic (TAfECG) acquisition of fetal heart rate (FHR) signals has recently been introduced into leading commercial cardiotocographic (CTG) monitors. Continuous wireless transmission of signals has raised the possibility of the technology being used during maternal mobilization in labor. This study aims to evaluate signal quality and accuracy of TAfECG acquisition of FHR signals during static and active maternal positions in labor when compared with Doppler signals and with the gold-standard method of fetal scalp electrode (FSE). MATERIAL AND METHODS: A total of 76 women with singleton term pregnancies in the active first stage of labor had simultaneously acquired FHR with TAfECG, Doppler, and FSE. Participants were asked to complete a supervised mobilization scheme, comprising five sequential 10-min periods of lying down, standing, sitting, walking, and rocking on the birthing ball. The three FHR signals were compared, defining signal loss as the percentage of signals under 20 bpm or exceeding 250 bpm and accuracy as the difference with FSE values. Computer analysis was used to quantify variability, accelerations, and decelerations. Static labor positions (lying down, standing, and sitting) were compared with active labor positions (walking and rocking on the birthing ball). RESULTS: Average signal loss was 5.3% with TAfECG (3.2% in static and 7.4% in active positions) and 15.5% with Doppler (8.3% in static and 30.7% in active positions). Average accuracy was 3.5 bpm with TAfECG (1.9 bpm in static and 5.04 bpm in active positions) and 13.9 bpm with Doppler (3.2 bpm in static and 24.7 bpm in active positions). Average variability was similar with TAfECG and FSE in static positions but significantly higher with TAfECG in active positions (23.6 vs. 13.5 bpm, p < 0.001). CONCLUSIONS: In static labor positions, TAfECG provides a low signal loss, similar to that obtained with FSE, and a good signal accuracy, so the technique can be considered reliable when the mother is lying down, standing, or sitting. During maternal movement, TAfECG causes an artificial increase in FHR variability, which can cause false reassurance regarding fetal oxygenation. Doppler signals are unreliable during maternal movements.
No takes yet. Share an insight, caveat, or question.
Reis‐de‐Carvalho et al. (2022) conducted an observational in Singleton term pregnancies in the active first stage of labor (n=76). Transabdominal electrocardiographic (TAfECG) monitoring vs. Doppler and fetal scalp electrode (FSE) was evaluated on Average signal loss. Transabdominal ECG provided lower average signal loss than Doppler (5.3% vs 15.5%) but artificially increased fetal heart rate variability during active maternal positions (23.6 vs 13.5 bpm, P<0.001).