Answers are grounded in this paper and cite their sources.
New Plastic membrane ion-selective electrode for buspirone hydrochloride based on buspironium tetraphenylborate was prepared. The electrode exhibited mean slope of calibration graph of 58.4 mV per decade of BusCl concentration at 25 degrees C. The electrode can be used within the concentration range 6.3 x 10(-5) - 10(-2) M BusCl at a pH range of 2.5-7.0. The standard electrode potentials were determined at different temperatures and used to calculate the isothermal temperature coefficient of the electrode, amounting to 0.00056 V degrees C(-1). The electrode showed a very good selectivity for BusCl with respect to a number of inorganic cations, sugars and amino acids. The electrode was applied to the potentiometric determination of the buspirone ion and its pharmaceutical preparation under batch and flow injection conditions. Also, buspirone was determined by conductimetric titrations. Graphite rod, copper and silver coated wire electrodes were prepared and characterized as sensors for the drug under investigation.
No takes yet. Share an insight, caveat, or question.
Abdel‐Ghani et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: