BACKGROUND: Our goal was to develop a suitably sensitive assay for N-acetyl-beta-D-hexosaminidase (HEX) and beta-glucuronidase to allow their use as markers of joint diseases. METHODS: We optimized a spectrophotometric method for the determination of lysosomally derived HEX and beta-glucuronidase in synovial fluid on a microplate reader to improve its utility. HEX and beta-glucuronidase act on the 4-nitrophenyl derivatives N-acetyl-beta-glucosamine and beta-D-glucuronide, respectively, to produce 4-nitrophenol, which can be measured at 405 nm on a microplate reader. RESULTS: Maximum enzyme activity was observed at pH 4.7 in a citrate-phosphate buffer for HEX and at pH 4.5 in an acetate buffer for beta-glucuronidase. A 10-microL sample with 30 microL of substrate solution and 40 microL of appropriate buffer produced measurable amounts of 4-nitrophenol after incubation for 60 min at 37 degrees Celsius. Reactions were terminated by the addition of 200 microL of 200 mM borate buffer (pH 9.8). CONCLUSIONS: The assay is sufficiently sensitive for small volumes of synovial fluid, and is useful for the clinical diagnosis of joint diseases.