PURPOSE: To develop an accelerated cardiac perfusion pulse sequence and test whether it is capable of increasing spatial coverage, generating high-quality images, and enabling quantification of myocardial blood flow (MBF). METHODS: We implemented an accelerated first-pass cardiac perfusion pulse sequence by combining radial k-space sampling, compressed sensing (CS), and k-space weighted image contrast (KWIC) filtering. The proposed and clinical standard pulse sequences were evaluated in a randomized order in 13 patients at rest. For visual analysis, 3 readers graded the conspicuity of wall enhancement, artifact, and noise level on a 5-point Likert scale (overall score index = sum of 3 individual scores). Resting MBF was calculated using a Fermi function model with and without KWIC filtering. Mean visual scores and MBF values were compared between sequences using appropriate statistical tests. RESULTS: , 128 ms, respectively). The overall image score index between accelerated (11.1 ± 1.3) and clinical standard (11.2 ± 1.3) was not significantly different (P = 0.64). Mean resting MBF values with KWIC filtering (0.9-1.2 mL/g/min across different slices) were significantly lower (P < 0.0001) than those without KWIC filtering (3.1-4.3 mL/g/min) and agreed better with values reported in literature. CONCLUSION: An accelerated, first-pass cardiac perfusion pulse sequence with radial k-space sampling, CS, and KWIC filtering is capable of increasing spatial coverage, generating high-quality images, and enabling quantification of MBF.
No takes yet. Share an insight, caveat, or question.
Naresh et al. (2018) studied this question.