Answers are grounded in this paper and cite their sources.
Based on measured R–T data for ten tempered Allen–Bradley resistors (BB, (1)/(8) W, 1 kΩ), a four-point-calibration scheme is described that yields ΔT/T≤0.5% (≤0.8%, 55–65 K) in the range 4.2–100 K. The procedure involves three successive approximations to the equation log R=A+BT−P.
No takes yet. Share an insight, caveat, or question.
Lawless et al. (1988) studied this question.