Disruption of polyurethane bipolar lead insulation by silk ligatures resulted in unipolar pacing and self-inhibition of ventricular output in a DVI pulse generator.
Case Report (n=1)
Direct application of suture material to polyurethane pacemaker leads should be avoided to prevent insulation disruption and subsequent pacemaker malfunction.
Direct silk ligation of polyurethane leads risks insulation failure and malfunction; this case report leaves open broader preventive recommendations.
Self-inhibition of bipolar DVI AV-sequential pacemakers has been attributed to electrode malposition, although the potential for self-inhibition due to current leakage has been recognized. Previous cases of current leakages due to insulation defects in bipolar VVI ventricular demand systems have usually presented as abnormal sensing or pacing function without a change to actual unipolar sensing and pacing. This article describes a case of disruption of the insulation of polyurethane bipolar atrial and ventricular leads by silk ligatures, resulting in unipolar atrial and ventricular pacing and self-inhibition of ventricular output of a DVI pulse generator. This case emphasizes the importance of avoiding any application of suture material directly to polyurethane leads. This case also emphasizes the importance of monitoring pacemaker artifact amplitude and axis in patient follow-up.
No takes yet. Share an insight, caveat, or question.
C. F. Sanford (1983) conducted a case report in Pacemaker malfunction (self-inhibition of DVI pulse generator) (n=1). Silk ligatures on polyurethane bipolar leads was evaluated on Self-inhibition of ventricular output. Disruption of polyurethane bipolar lead insulation by silk ligatures resulted in unipolar pacing and self-inhibition of ventricular output in a DVI pulse generator.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: