Key result
Repair of an epicardial electrode insulation defect restored normal pacemaker function by preventing myopotential leakage that had suppressed the bipolar demand unit.
Case Report (n=1)
An electrode insulation defect can allow myopotentials to leak into a bipolar pacing system and suppress a demand pacemaker, which can be corrected by repairing the defect.
Answers are grounded in this paper and cite their sources.
Identifies correctable insulation defect causing myopotential pacemaker suppression; single case leaves open broader recognition and screening.
A patient is described in whom myopotentials orginating from the anterior abdominal wall muscle suppressed the implanted demand pacemaker despite its bipolar mode of action. This phenomenon was shown by simultaneous recording of the electrocardiogram the electromyogram. At operation, a defect in the insulation of a previously repaired epicardial electrode was found lying in close proximity to these muscles. After repair of the insulation defect, normal pacemaker function was restored. It is suggested that the myopotentials leaked into the pacing system through the insulation defect, thereby suppressing the demand unit, which maintained its bipolar mode of pacing throughout.
No takes yet. Share an insight, caveat, or question.
Amikam et al. (1977) conducted a case report in Pacemaker suppression by myopotentials (n=1). Repair of electrode insulation defect was evaluated on Restoration of normal pacemaker function. Repair of an epicardial electrode insulation defect restored normal pacemaker function by preventing myopotential leakage that had suppressed the bipolar demand unit.