Key result
Diaphragmatic myopotentials caused repeated transient suppression of a normally functioning bipolar demand pulse generator during maneuvers such as deep respiration, straining, and coughing.
Case Report
Sensing of diaphragmatic myopotentials should be considered in the differential diagnosis of unexplained pacemaker pauses in patients with bipolar demand pacemakers.
Answers are grounded in this paper and cite their sources.
Diaphragmatic myopotentials may cause pauses in bipolar pacemakers; this case report leaves open prevalence and management in contemporary devices.
This report describes inhibition of a normally functioning bipolar demand pulse generator by diaphragmatic myopotentials. Transient pacemaker suppression occurred repeatedly with deep respiration, straining, the Valsalva maneuver, coughing, sneezing and laughing. When the magnet was applied, none of these maneuvers inhibited the pacemaker. Extensive investigations ruled out an intermittent electrode problem such as a wire fracture or insulation break. Sensing of diaphragmatic myopotentials should be considered in the differential diagnosis of unexplained pacemaker pauses.
No takes yet. Share an insight, caveat, or question.
Barold et al. (1977) conducted a case report in Pacemaker inhibition. Diaphragmatic myopotentials was evaluated on Transient pacemaker suppression. Diaphragmatic myopotentials caused repeated transient suppression of a normally functioning bipolar demand pulse generator during maneuvers such as deep respiration, straining, and coughing.