Deep inspiration caused intermittent failure of pacing in a patient with a permanent demand pacemaker, likely due to inhibition by non-cardiac potentials associated with respiration.
Case Report (n=1)
Highlights the potential for myopotentials associated with respiration to inhibit demand pacemakers with unipolar electrode systems.
Answers are grounded in this paper and cite their sources.
May warrant checking for respiratory myopotential inhibition in symptomatic unipolar pacemaker patients; hypothesis-generating case that leaves open modern device implications.
A 69-year-old man with a permanent demand pacemaker and a unipolar electrode system had intermittent failure of pacing during deep inspiration. Pacing was not interrupted when the unit was switched to fixed rate mode by an external magnet. Thus, the problem was not caused by a change in electrode position or to a loose connexion. The likely cause was inhibition by non-cardiac potentials, possibly myopotentials associated with respiration.
No takes yet. Share an insight, caveat, or question.
Peter et al. (1976) conducted a case report in Intermittent failure of pacing (n=1). Deep inspiration was evaluated on Intermittent failure of pacing. Deep inspiration caused intermittent failure of pacing in a patient with a permanent demand pacemaker, likely due to inhibition by non-cardiac potentials associated with respiration.