Key result
Sudden changes of resistance in the electrode-lead system due to intermittent fractures can produce potential changes interpreted as spontaneous ventricular depolarization, falsely inhibiting demand pacemakers.
Intermittent lead fractures can cause sudden resistance changes that mimic spontaneous ventricular depolarization, leading to false inhibition of demand pacemakers and presenting a diagnostic challenge.
Answers are grounded in this paper and cite their sources.
Clinicians should suspect intermittent lead fracture in unexplained demand pacemaker inhibition; leaves open need for modern detection algorithms.
Certain variations in discharge rate of demand pacemakers may be associated with intermittent fractures and resulting resistance changes in the electrode-lead system. These variations in rate are related to inhibition of the demand circuit of the pacemaker. Most pacemakers produce a biphasic output pulse which results in a low amplitude current during the interval between stimuli. Therefore, sudden changes of resistance in the electrode-lead system will produce a potential change at the pacemaker terminals which is of sufficient magnitude to be interpreted as spontaneous ventricular depolarization and inhibit the demand pacemaker. Such inhibition may occur in the absence of other signs of lead fracture and present a difficult differential diagnostic problem of pacemaker malfunction.
No takes yet. Share an insight, caveat, or question.
Lasseter et al. (1970) studied Demand pacemaker malfunction. Intermittent fractures and resistance changes in electrode-lead system was evaluated on False inhibition of demand pacemakers. Sudden changes of resistance in the electrode-lead system due to intermittent fractures can produce potential changes interpreted as spontaneous ventricular depolarization, falsely inhibiting demand pacemakers.