Key result
Intraoperative recordings of the ventricular electrogram in 11 patients with demand pacemakers demonstrated spurious signals mimicking an R-wave, indicating intermittent pacing lead fracture.
Case Report (n=11)
Pacing and recording on the same lead is an effective technique for detecting spurious signals indicative of intermittent pacing lead fracture in patients with demand pacemakers.
Answers are grounded in this paper and cite their sources.
May aid intraoperative detection of intermittent lead fracture; leaves open prospective validation before wider adoption.
This report describes some of the various techniques which may be used to demonstrate intermittent and/or incomplete fracture of a pacing lead. Eleven patients treated with an R-wave-inhibited demand pacer showed signs of paroxysmal arrhythmias. Intraoperative recordings of the ventricular electrogram showed the presence of spurious signals mimicking an R-wave. These signals varied from patient to patient but remained morphologically constant when a given technique was used in a given case. The most significant signals were picked up when pacing and recording on the same lead.
No takes yet. Share an insight, caveat, or question.
Múgica et al. (1978) conducted a case report in Demand pacemaker arrhythmias due to pacing lead fracture (n=11). Intraoperative recordings of the ventricular electrogram was evaluated on Presence of spurious signals mimicking an R-wave. Intraoperative recordings of the ventricular electrogram in 11 patients with demand pacemakers demonstrated spurious signals mimicking an R-wave, indicating intermittent pacing lead fracture.