Key result
Recording high voltage unipolar electrograms during bipolar demand ventricular pacing does not guarantee adequate QRS signal delivery to the pacemaker, which can lead to pacemaker failure.
Case Report (n=2)
In bipolar pacing systems, adequate QRS signal delivery depends on the potential difference between the two intracardiac electrodes, which should be measured directly.
Answers are grounded in this paper and cite their sources.
May warrant direct inter-electrode potential measurement in bipolar pacing; leaves open whether unipolar signals reliably predict delivery.
The recording of high voltage unipolar electrograms during bipolar demand ventricular pacing does not always guarantee delivery of an adequate QRS signal to the pacemaker as illustrated in two cases described in this report. In a bipolar system the actual QRS signal delivered to the pacemaker is the potential difference between the two intracardiac electrodes and may be measured by recording bipolar electrograms on a standard electrocardiograph.
No takes yet. Share an insight, caveat, or question.
S. Serge Barold (1971) conducted a case report in Demand pacemaker failure (n=2). Bipolar demand ventricular pacing was evaluated on Demand pacemaker failure. Recording high voltage unipolar electrograms during bipolar demand ventricular pacing does not guarantee adequate QRS signal delivery to the pacemaker, which can lead to pacemaker failure.