Key result
Myopotential inhibition occurred in 47% of patients with unipolar pacemakers during reach maneuvers, with susceptibility varying from 33% to 78% depending on the manufacturer.
Why the study?
Do unipolar pacemakers with advanced sensing amplifiers and insulative coating reduce the incidence of myopotential inhibition compared to standard models?
Observational (n=77)
Do unipolar pacemakers with advanced sensing amplifiers and insulative coating reduce the incidence of myopotential inhibition compared to standard models?
Advanced sensing amplifiers and insulative coating in unipolar pacemakers significantly reduce the incidence and severity of myopotential inhibition during physical maneuvers.
Answers are grounded in this paper and cite their sources.
Myopotential inhibition remains frequent in unipolar pacemakers; manufacturer variation supports hypothesis that advanced amplifiers may reduce risk but requires prospective confirmation.
Myopotential oversensing by unipolar pulse generators can cause patient symptoms ranging from dizziness and syncope to ventricular tachycardia. Seventy-seven patients with implanted unipolar pacemakers from three manufacturers participated in isometric and reach exercises to evaluate their pacemakers' susceptibility to myopotentials. Myopotential inhibition occurred in 47% of the patients performing the reach maneuver. Testing revealed a wide difference in level of susceptibility to myopotentials between pacemakers of different manufacture (a low of 33% inhibition for the least susceptible to a high of 78% inhibition for the most susceptible during the reach maneuver). The normal pacing interval was extended by myopotential oversensing for each manufacturer's model within a range of 0.3-3.9 seconds. Pulse generators incorporating additional automatically-adjusting threshold and reversion circuits in the sense amplifier along with standard bandpass filtering exhibited: a) two-to-three times less susceptibility to myopotentials; and b) a 75% reduction in the maximum pacing interval extension as compared with pacemakers with bandpass filtering alone. The effectiveness of insulative coating in reducing myopotential inhibition was substantiated as coated pulse generators had a 22% lower incidence of muscle sensing than those that were uncoated. Six out of seven patients tested had symptoms during Holter monitoring which correlated with pacemaker myopotential inhibition. Selecting pulse generators with improved sensing amplifiers, clinical testing of patients with unipolar pacemakers using the reach method, and reprogramming of sensitivity will significantly reduce the incidence of myopotential inhibition.
No takes yet. Share an insight, caveat, or question.
Fetter et al. (1984) conducted an observational in Implanted unipolar pacemakers (n=77). Unipolar pacemakers during isometric and reach exercises was evaluated on Myopotential inhibition during the reach maneuver. Myopotential inhibition occurred in 47% of patients with unipolar pacemakers during reach maneuvers, with susceptibility varying from 33% to 78% depending on the manufacturer.