Key result
Electromyopotential oversensing in unipolar pacemakers was frequently provoked, occurring in 61% of patients for atrial sensing and 39% for ventricular sensing.
Why the study?
What is the frequency of electromyopotential oversensing provocation in current unipolar pacemakers?
Observational (n=34)
What is the frequency of electromyopotential oversensing provocation in current unipolar pacemakers?
Electromyopotential oversensing remains common in modern unipolar pacemakers, particularly at high sensitivities used for atrial sensing, highlighting the need to avoid unnecessarily high sensitivity settings.
Frequent oversensing may support avoiding high atrial sensitivities in unipolar pacemakers; leaves open prospective outcome data.
UNLABELLED: Electromyopotential oversensing of unipolar pacemakers was first appreciated 20 years ago, but its prevalence in present day devices is less well defined. Thirty-four pacemaker patients, only two with symptoms suggestive of oversensing, were evaluated in unipolar settings to assess the frequency of provocation of oversensing in one or, if present, both pacing channels. The sensing threshold of each patient, whenever possible, was recorded as well. RESULTS: Atrial oversensing occurred in 11/18 patients (61%), all at sensitivities in the 0.4-1.0 mV range. Ventricular oversensing was noted in 13/33 patients (39%), with all but one programmed to settings of 1.25 mV or more sensitive (i.e. < 1.25 mV). Twenty six of 26 patients amenable to testing had ventricular sensing thresholds of at least 4.0 mV or more. Of the 15 patients amenable to atrial sensing threshold testing, 4 had a threshold of 1.0 mV or < 1.0 mV, 6 had thresholds between 1.0-2.0 mV, and 5 sensed at settings > 2.0 mV. CONCLUSION: Electromyopotential oversensing remains a relevant issue in current day unipolar pacemakers. Most patients do not describe symptoms related to electromyopotential interference, yet such interference is frequently provoked. Oversensing is common at high sensitivities typically utilized for atrial sensing, but quite unusual at settings necessary for adequate ventricular sensing. Programming unipolar devices to unnecessarily high sensitivities should be avoided or serious consequences may result.
No takes yet. Share an insight, caveat, or question.
Gross et al. (1992) conducted an observational in Pacemaker patients (n=34). Unipolar pacemaker settings was evaluated on Frequency of provocation of electromyopotential oversensing. Electromyopotential oversensing in unipolar pacemakers was frequently provoked, occurring in 61% of patients for atrial sensing and 39% for ventricular sensing.