Key result
A numerical model of the PUCA pump determined that a catheter internal diameter of at least 6.95 mm is required for a flow of 5 L/min, and 5.50 mm for 3 L/min, which was validated in vitro.
A numerical model successfully determined the required catheter diameters for the PUCA pump to achieve target blood flows, which was validated by in vitro testing.
Answers are grounded in this paper and cite their sources.
Guides minimum catheter diameters for target PUCA flows; leaves open in vivo performance and clinical translation.
The pulsating catheter (PUCA) pump, a left ventricular assist device, consists of a hydraulically or pneumatically driven membrane pump, extracorporeally placed and mounted to a valved catheter. The catheter is introduced into an easily accessible artery and positioned with its distal tip in the left ventricle. Blood is aspirated from the left ventricle during systole and ejected into the ascending aorta during diastole. A numerical model of the PUCA pump has been developed to determine the internal diameter of the PUCA pump catheter that allows a certain blood flow. The model considers a limitation of mechanical blood damage and determines the accompanying pressure and flow profile for driving the pump. For a flow of 5 L/min, a catheter with an internal diameter of at least 6. 95 mm is required. For 3 L/min, the minimal diameter is 5.50 mm. The latter catheter can be introduced in the axillary artery, the former via the aorta during an open thorax surgical procedure. To validate the numerical model, 2 different PUCA pump configurations were tested in vitro. Results showed a good resemblance between model and in vitro behavior of the PUCA pump.
No takes yet. Share an insight, caveat, or question.
Mihaylov et al. (1999) studied this question. Pulsating catheter (PUCA) pump was evaluated on Internal diameter of the PUCA pump catheter required for a certain blood flow. A numerical model of the PUCA pump determined that a catheter internal diameter of at least 6.95 mm is required for a flow of 5 L/min, and 5.50 mm for 3 L/min, which was validated in vitro.
Synapse has enriched 10 related papers. Consider them for comparative context: