Answers are grounded in this paper and cite their sources.
A diffractive zone plate provides a highly linear wavelength-to-depth coding, allowing for nonmechanical depth scanning in a confocal microscope. This chromatic confocal microscope, constructed with 40x and 60x objectives, achieves axial position changes of 55 and 25 mum, respectively, for a wavelength tuning range of 100 nm. The corresponding longitudinal point-spread functions are measured and shown to possess full-width half-maximums of 2.52 and 2.23 mum, respectively. Two-dimensional profiles of a two-phase-level grating and a four-phase-level diffractive structure are given. The performance of the chromatic confocal microscope is consistent with that of the conventional confocal operation of the microscope.
No takes yet. Share an insight, caveat, or question.
Dobson et al. (1997) studied this question.