Answers are grounded in this paper and cite their sources.
The probe was designed for the measurement of DC voltages of up to 30 kV from high impedance sources. It is based on a resistive divider with a total resistance of 200 GΩ and a step-down factor of 10’000. In order to allow the measurement of the stepped down voltage with a conventional multimeter without loading, the signal was buffered with an operational amplifier. The device was calibrated against a commercial probe using a low impedance high voltage source. A linear relationship was obtained for a high impedance resistive ladder for voltages between 3 and 30 kV, with a coefficient of determination (R2) of 0.9999. The low-cost device (ca. US $200) fills an application niche not addressed by commercial products.
No takes yet. Share an insight, caveat, or question.
Boillat et al. (2024) studied this question.