Answers are grounded in this paper and cite their sources.
Titanium lumbar interbody spinal fusion devices are reported to be 90% effective for single-level lumbar interbody fusion, although radiographic determination of fusion has been debated. Using blinded radiographic, biomechanic, histologic, and statistical measures, researchers in the present study evaluated a radiolucent 70/30 poly(L-lactide-co-D,L-lactide) (70/30 D,L-PLa) interbody fusion device packed with autograft or rhBMP-2 on a collagen sponge in 25 sheep at 3, 6, 12, 18, and 24 months. A trend of increased fusion stiffness, radiographic fusion, and histologic fusion was demonstrated from 3 months to 24 months. Device degradation was associated with a mild to moderate chronic inflammatory response at all postoperative sacrifice times.
No takes yet. Share an insight, caveat, or question.
Toth et al. (2002) studied this question.