Key result
Optical techniques for measuring AAV full-to-total ratio showed <10% SD between laboratories, whereas AUC had ≈25% SD and PCR-ELISA demonstrated poor reproducibility.
Why the study?
Therapeutic safety and efficacy of recombinant AAV depend on correct DNA encapsidation characterized by the full-to-total ratio, but interlaboratory and intertechnique variations required evaluation.
Optical techniques provide the most reproducible interlaboratory measurements for AAV full-to-total capsid ratios, whereas PCR/ELISA and AUC require further standardization.
Answers are grounded in this paper and cite their sources.
Optical methods enable reliable AAV capsid quantification across labs; leaves open standardization of PCR/ELISA and AUC for clinical vector use.
Recombinant adeno-associated virus (AAV) is one of the main viral vector-based gene therapy platforms. AAV is a virus consisting of a ≈25 nm diameter capsid with a ≈4.7 kb cargo capacity. Therapeutic safety and efficacy depend on the correct encapsidation of the DNA in individual virus particles, which is often characterized by the single scalar value of the ratio of full capsids with complete viral genomes to the total viral capsid number [the full-to-total (FTT) ratio]. This study reports on the interlaboratory and intertechnique variations of measurement methods for FTT among a cohort of organizations. The analytical methods used were sedimentation velocity analytical ultracentrifugation (SV-AUC) with UV/Vis and/or Rayleigh interference optics, size exclusion chromatography (SEC) with multi-angle light scattering (MALS), and tandem UV/Vis and/or refractive index, cryogenic electron microscopy, dual-wavelength ultraviolet spectrophotometry, and ELISA coupled with quantitative PCR (qPCR, dPCR, or ddPCR). FTT measurements for both AAV5 and AAV8 serotypes were similar, except for PCR-ELISA. The optical techniques (UV spectroscopy/SEC-MALS) showed <10% SD between laboratories, likely from the uniformity of existing industry protocols. AUC, while demonstrating good repeatability, had ≈25% SD interlaboratory, suggesting the need for standardized methods. PCR and ELISA had poor reproducibility due to variations in both PCR and ELISA protocols and instrumentation. The discussion presents intended future efforts to improve and harmonize these measurements to increase both the repeatability and reproducibility of AAV viral particle critical quality attributes such as FTT.
No takes yet. Share an insight, caveat, or question.
Lehman et al. (2024) studied Adeno-Associated Virus (AAV) vector characterization. Measurement methods for full-to-total (FTT) ratio was evaluated on Interlaboratory and intertechnique variations of measurement methods for FTT. Optical techniques for measuring AAV full-to-total ratio showed <10% SD between laboratories, whereas AUC had ≈25% SD and PCR-ELISA demonstrated poor reproducibility.
Synapse has enriched 10 related papers. Consider them for comparative context: