Key result
Heparin clearance during extracorporeal circulation in infants was virtually identical to the sum of clearances in the patient and isolated circuit.
Why the study?
How is heparin cleared during extracorporeal membrane oxygenation in infants?
Observational (n=5)
How is heparin cleared during extracorporeal membrane oxygenation in infants?
More than half of the heparin administered to infants during ECMO is eliminated by the extracorporeal circuit itself, explaining the relatively large doses needed to maintain anticoagulation.
Answers are grounded in this paper and cite their sources.
May explain higher heparin needs in infant ECMO; extends clearance models but leaves dosing optimization open.
Heparin anticoagulation is necessary to prevent clotting during procedures involving the extracorporeal circulation of blood. Our preliminary observations suggested that heparin was inactivated in the extracorporeal circuit during extracorporeal membrane oxygenation. We tested this hypothesis by comparing heparin pharmacokinetics in five infants during extracorporeal circulation with kinetics, respectively determined in each patient and in the isolated circuit immediately after discontinuation of the procedure. Heparin clearance was 1.6 +/- 0.5 ml/kg/min in the patient and 2.1 +/- 0.8 ml/kg/min in the separated circuit. In each patient, the total of heparin clearances in the patient and circuit, 3.7 +/- 1.0 ml/kg/min, was virtually identical with the heparin clearance during the procedure, 3.8 +/- 1.9 ml/kg/min (r = 0.94, p less than 0.01). We conclude that more than one half of the heparin administered to infants during extracorporeal membrane oxygenation is eliminated by the extracorporeal circuit itself or by blood components in the circuit. These data explain the relatively large heparin doses needed to maintain anticoagulation in infants during extracorporeal circulation. In light of these findings, a reexamination of the normal mechanisms of elimination of heparin activity appears to be warranted.
No takes yet. Share an insight, caveat, or question.
Green et al. (1990) conducted an observational in Infants requiring extracorporeal membrane oxygenation (n=5). Extracorporeal membrane oxygenation vs. Isolated circuit and patient after discontinuation was evaluated on Heparin clearance (r = 0.94, p=<0.01). Heparin clearance during extracorporeal circulation in infants was virtually identical to the sum of clearances in the patient and isolated circuit.