This study introduces an improved method for the on-site quantification of cadmium and lead in officinal plants using differential pulse anodic stripping voltammetry (DP-ASV) with a glassy carbon electrode (GCE) modified by an in-situ mercury film (iMF-GCE). Initial experiments, conducted with one-variable-at-the-time (OVAT) approach, have provided suboptimal recovery rates and linearity, factors refined later through a systematic experimental approach involving two different designs: Plackett-Burman and Face Centered Composite Design (FCCD). Certified reference materials confirmed the method's reliability and its potential for effectiveness. The key parameters were optimized, achieving an Edep of -1.20 V and a tdep of 195 s for optimal performance. The results obtained allowed substantial improvements in accuracy, with recovery rates of 85.8% for Cd and 96.4% for Pb, and significantly lowered detection limits from, 1.54 μg L -1 to 0.63 μg L -1 for Cd and from 0.15 μg L -1 to 0.045 μg L -1 for Pb. Validation through field applications demonstrated the method's suitability for the analysis of official plants. Other applications are for environmental monitoring and food safety control. • The optimization improved Sensitivity and Accuracy of the method thakns to DOE application. • The setup is simple, requires minimal reagents, and can be deployed in the field. • Plant results obtained on-site showed good agreement with laboratory results. • This method could be used also for environmental monitoring and food analysis.
No takes yet. Share an insight, caveat, or question.
Favilli et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: