Key result
The PLx Multi-Code Respiratory Virus Panel demonstrated high accuracy for detecting respiratory viruses, with sensitivities of 94-100% and specificities of 99-100% for individual viruses.
Why the study?
Does the PLx-RVP assay improve the detection of respiratory viruses compared to conventional virologic testing in archived respiratory specimens?
Observational (n=410)
Does the PLx-RVP assay improve the detection of respiratory viruses compared to conventional virologic testing in archived respiratory specimens?
The PLx-RVP is a highly accurate multiplex assay that significantly improves the detection rate of respiratory viruses compared to conventional testing.
Answers are grounded in this paper and cite their sources.
Supports multiplex respiratory virus detection in research; leaves open prospective clinical validation before routine use.
High-throughput multiplex assays for respiratory viruses are an important step forward in diagnostic virology. We compared one such assay, the PLx Multi-Code Respiratory Virus Panel (PLx-RVP), manufactured by Eragen Biosciences, Inc. (Madison, WI), with conventional virologic testing, consisting of fluorescent-antibody staining plus testing with the R-mix system and fibroblast tube cultures. The test set consisted of 410 archived respiratory specimens, mostly nasopharyngeal swabs, including 210 that had been positive by conventional testing for a balanced selection of common respiratory viruses. Specimens yielding discrepant results were evaluated using a panel of respiratory virus PCR assays developed, characterized, and validated with clinical specimens. PLx-RVP increased the total rate of detection of viruses by 35.8%, and there was a 25.7% increase in the rate of detection of positive specimens. Reference PCR assay results corroborated the PLx-RVP result for 54 (82%) of 66 discrepancies with conventional testing. Of the 12 specimens with discrepancies between PLx-RVp and the reference PCRs, 6 were positive for rhinovirus by PLx-RVP and the presence of rhinovirus was confirmed by nucleotide sequencing. The remaining six specimens included five in which the PLx-RVP failed to detect parainfluenza virus and one in which the detection of influenza A virus by PLx-RVP could not be confirmed by the reference PCR. Taking the results of the reference PCR assay results into account, the sensitivities of the PLx-RVP for individual viruses ranged from 94 to 100% and the specificities ranged from 99 to 100%. We conclude that PLx-RVP is a highly accurate system for the detection of respiratory viruses and significantly improves the rate of detection of these viruses compared to that by conventional virologic testing.
No takes yet. Share an insight, caveat, or question.
Arens et al. (2010) conducted an observational in Respiratory viruses (n=410). PLx Multi-Code Respiratory Virus Panel (PLx-RVP) vs. Conventional virologic testing and reference PCR was evaluated on Detection of respiratory viruses (sensitivity and specificity) (Sensitivities 94-100%, specificities 99-100%). The PLx Multi-Code Respiratory Virus Panel demonstrated high accuracy for detecting respiratory viruses, with sensitivities of 94-100% and specificities of 99-100% for individual viruses.