Answers are grounded in this paper and cite their sources.
Experimental study demonstrates a stable two-micrometer supercontinuum laser yielding high axial resolution in optical coherence tomography, indicating strong utility in highly scattering media.
We report a compact and stable broadband light source for 2-µm optical coherence tomography (OCT). The source is seeded by a polarization-maintaining (PM)-based mode-locked thulium-doped fiber oscillator with an intracavity spectral filter that suppresses soliton sidebands and provides a stable seed for subsequent nonlinear conversion. After wavelength shifting in a thulium-doped fiber amplifier and spectral broadening in a 5-m highly nonlinear fiber, a supercontinuum (SC) spectrum with a 295-nm 3-dB bandwidth was obtained in the 2-µm region. A fiber-based time-domain OCT system was employed as a broadband characterization platform, yielding an axial resolution of 6.6 µm and a sensitivity of 99.6 dB. Representative OCT imaging results further confirmed the practical performance of the developed 2-µm system. These results demonstrate a practical route toward stable 2-µm SC generation for OCT imaging in highly scattering media.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: