Key result
The Bard PowerFlow Implantable Apheresis IV Port was accessed on the first attempt in all 47 procedures across 4 patients, with pressure alarms occurring in only 6 procedures.
Why the study?
Does the Bard PowerFlow Implantable Apheresis IV Port provide adequate flow rates and reliable access for patients undergoing therapeutic plasma exchange?
Observational (n=4)
Does the Bard PowerFlow Implantable Apheresis IV Port provide adequate flow rates and reliable access for patients undergoing therapeutic plasma exchange?
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
The PowerFlow implantable venous access device demonstrates promising performance and reliable access for patients requiring therapeutic plasma exchange.
We prospectively evaluated the Bard PowerFlow Implantable Apheresis IV Port in four patients undergoing outpatient therapeutic plasma exchange over 18 to 97 days. Three had bilateral internal jugular access ports, and one had a single left internal jugular access port for the inlet line with return via antecubital vein. Two patients receiving 5% albumin as replacement fluid achieved peak inlet flow of 99 ± 5 mL/min and 101 ± 6 mL/min, and peak plasma flow of 53 ± 6 and 47 ± 6 mL/min, respectively. Two patients receiving plasma as replacement fluid achieved peak inlet flow of 46 ± 7 and 85 ± 4 mL/min and peak plasma flow of 27 ± 3 and 35 ± 4 mL/min, respectively. Apheresis nurses accessed the ports on the first attempt in all procedures. Pressure alarms occurred in 6 of 47 procedures and were easily resolved by lowering the inlet rate by 10% in 5 of them. The PowerFlow shows promise as an implantable venous access device for apheresis.
Garrity et al. (2019) conducted an observational in Outpatient therapeutic plasma exchange (n=4). Bard PowerFlow Implantable Apheresis IV Port was evaluated on Performance characteristics including peak inlet flow, peak plasma flow, and access success. The Bard PowerFlow Implantable Apheresis IV Port was accessed on the first attempt in all 47 procedures across 4 patients, with pressure alarms occurring in only 6 procedures.