Key result
Covered stent implantation cuts thoracic aortic trans-stenotic gradient by ~41 mmHg in pediatric coarctation.
Why the study?
Treating severe descending thoracic aorta stenosis due to Takayasu arteritis in a small child with acute left heart failure and small vascular access presents significant interventional challenges.
Case Report (n=1)
No
A coaxial 8Fr-in-10Fr sheath technique with suture-mediated vascular closure enables safe and effective covered stent implantation for severe descending thoracic aortic stenosis in a small child (25 kg) with small femoral arteries.
Answers are grounded in this paper and cite their sources.
Supports feasibility of coaxial sheath technique in small children; hypothesis-generating and should not yet change practice.
A 10-year-old girl (23 kg) having a medical history of uncontrolled hypertension was presented to our hospital because of acute left heart failure. Transthoracic echocardiography showed stenosis of descending thoracic aorta with a maximum trans-stenotic pressure gradient of 50 mmHg and severe left ventricular systolic dysfunction with an ejection fraction of 20%. She was diagnosed with Takayasu arteritis with a long severe stenosis of segment III of the thoracic aorta. The procedure of percutaneous transluminal angioplasty was performed and helped to reduce the pressure gradient significantly. After a 6-month follow-up, the left ventricular function was unimproved. Hence, aortic angiography was done and revealed the descending thoracic aorta restenosis with a pressure gradient of 46 mmHg. Despite the difficulties of small vascular access and the disease severity, this patient was intervened by cover stent without any complications. The trans-stenotic pressure gradient decreased remarkably to 5 mmHg. The stent implantation should be considered in the severe stenosis of descending thoracic aorta because of its benefit and safety.
No takes yet. Share an insight, caveat, or question.
Ho et al. (2021) conducted a case report in Severe descending thoracic aorta stenosis (Takayasu arteritis / atypical coarctation) with acute left heart failure (n=1). Percutaneous transluminal angioplasty followed at 6 months by balloon-expandable covered stent implantation (two 12x59 mm Begraft stents) using a coaxial sheath introduction technique (8Fr Amplatzer delivery inside 10Fr Cook outer sheath) was evaluated on Trans-stenotic pressure gradient across the descending thoracic aorta. Covered stent implantation via coaxial sheath technique reduced the descending thoracic aorta trans-stenotic gradient from 46 to 5 mmHg and increased LVEF by 25% in a 10-year-old girl.
Synapse has enriched 10 related papers. Consider them for comparative context: