An electrical impedance tomography (EIT) system has been constructed, operating at two frequencies, 40.96 and 81.92 kHz, for investigating the practicability of the dual-frequency imaging method discussed theoretically in a previous paper. For testing the system, a phantom with a frequency-dependent electrical conductivity was designed. The properties of the phantom can be adjusted to match the frequency dependence observed in a given type of tissue. Dual-frequency images were obtained from a phantom simulating liver and also from 200 g of porcine liver in a saline tank. Prior to image reconstruction, it was necessary to apply a correction to the data to cancel the effects of stray capacitance within the electronics.