A method for measuring both the real and imaginary parts of complex shear-wave impedance Z s * for low viscosity liquids using a planar AT-quartz resonator (thickness shear) in an RF bridge network between 2 and 34 MHz, is described. This technique employs a PLL circuit and permits continuous quantitative recording of the sample impedance (probe volume below 1 ml). The mechanical vibration amplitude at the transducer-liquid interface is in the Angstrom range, preventing any heating and structural damage effects. Incremental Z s -resolution is better than 50 kg m -2 s -1 ( approximately 1% of water impedance at 6 MHz), as is demonstrated by test measurements on different newtonian liquids. These measurements show that Z s * values determine quantitatively the shear quartz resonator parameters. This method permits measurements of interface and surface layer properties.
No takes yet. Share an insight, caveat, or question.
Eggers et al. (1987) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: