This is the authors' abstract. We don't add key points for this paper.
We have constructed a scanning tunneling microscope for operation under UHV conditions (10−8 Pa). With this instrument topographic measurements can be made on metal and semiconductor surfaces by means of a scanning tip electrode, driven by piezoelectric ceramic elements. The maximum area covered ranges up to 4000×4000 Å2 with a resolution better than 10 Å laterally and 0.15 Å rms perpendicular to the plane. Because of its compact design, the scan unit is very insensitive to vibrations and has a response time down to 0.3 ms. This allows a high scan rate to be used. In order to minimize temperature effects, special attention is paid to the geometry of the construction and the materials used, resulting in a drift ≤4 Å/min along the surface and 0.5 Å/min perpendicular.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Walle et al. (1985) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: