This is the authors' abstract. We don't add key points for this paper.
) were compared with reference values using Student's t-test and chi-square test. Results For anteversion, mean values for the PRV (27.34° ± 7.27°) and the new software (27.29° ± 7.21°) were not significantly different (p = 0.49). The new software differed from the PRV by a mean of 0.05° ± 0.93°. Similar results were noted for inclination, where the new software differed from the PRV and SRV by -0.13° ± 0.65° and 0.25° ± 1.26°, respectively (mean values: PRV: 43.62° ± 6.02°; SRV: 43.99° ± 6.27°; new software: 43.74° ± 6.17°; p = 0.87), and for leg length, where the new software differed from the PRV and SRV by 0.05 mm ± 0.46 mm and 0.22 mm ± 0.52 mm, respectively (mean values: PRV: 10.61 mm ± 11.60 mm; SRV: 10.77 mm ± 11.70 mm; new software: 10.56 mm - ± 11.61 mm; p = 0.98). Measurements were highly correlated across multiple reviewers (intraclass correlation coefficient ≥0.987). Conclusions The new software measurement tool is accurate and precise for assessing the acetabular component position and leg length measurements following THA in AP pelvic radiographs compared to currently used image measurement software.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Muir et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: