Key result
Engraving a simple code onto the hard plastic connector of laryngeal mask airways allowed tracking of sterilizations and prevented further incidents of broken or failed masks.
Engraving a code on the hard plastic connector of laryngeal mask airways allows for tracking of sterilization cycles and timely replacement, preventing equipment failure.
Answers are grounded in this paper and cite their sources.
May support sterilization tracking to avert laryngeal mask failures; leaves open generalizability beyond single-center experience.
The fractured laryngeal mask airway (LMA) reported by Drs Crawford and Davidson (Anaesthesiu 1992; 47: 76) is similar to an incident reported by me (letter submitted). We also found it impossible to mark the laryngeal masks (LM) indelibly with any ink or paint. However, we have been able to engrave a simple code onto the hard plastic connector at the proximal end of the tube (Fig. This is done using an industrial engraver (Powerline 172) and does not appear to damage the connector. The code is noted at each pass through the Central Sterile Supply Department and the LM replaced after 40 sterilisations. Since starting this routine we have had no further incidents with broken or failed LMs.
No takes yet. Share an insight, caveat, or question.
S.J. Squires (1992) reported a letter. Engraving a code onto the hard plastic connector was evaluated on Incidents with broken or failed laryngeal masks. Engraving a simple code onto the hard plastic connector of laryngeal mask airways allowed tracking of sterilizations and prevented further incidents of broken or failed masks.