Key result
A novel blood noise reduction algorithm improved intraobserver and interobserver correlation coefficients for lumen cross-sectional area in IVUS images.
Why the study?
Does a novel blood noise reduction algorithm improve the reproducibility of quantitative measurements in ultra-high-frequency IVUS images in patients with coronary artery disease?
Observational (n=35)
Does a novel blood noise reduction algorithm improve the reproducibility of quantitative measurements in ultra-high-frequency IVUS images in patients with coronary artery disease?
A novel blood noise reduction algorithm significantly improves the reproducibility of lumen and plaque measurements in ultra-high-frequency IVUS imaging.
Answers are grounded in this paper and cite their sources.
May improve ultra-high-frequency IVUS measurement reproducibility; hypothesis-generating and requires prospective validation before clinical adoption.
BACKGROUND: Ultra-high-frequency (40- to 50-MHz) intravascular ultrasound (IVUS) improves image quality compared with conventional 20- to 30-MHz IVUS. However, as the frequency of IVUS increases, high-intensity backscatter from blood components may cause visual difficulties in discrimination between the lumen and arterial wall structure. The purpose of this study was to evaluate the effect of a novel blood noise reduction algorithm (BNR) on quantitative coronary ultrasound measurements. METHODS AND RESULTS: IVUS studies using a 40-MHz transducer were performed in 35 patients with coronary artery disease. A total of 620 gray-scale images (310 pairs) were processed with and without the BNR, and lumen cross-sectional area (CSA) was determined by 2 independent observers. With the BNR, the intraobserver and interobserver correlation coefficients for lumen CSA were significantly improved (0.85 to 0.99 and 0.80 to 0.98, respectively). In the 270 images (135 pairs) in which vessel wall measurements were possible, the BNR significantly improved the intraobserver and interobserver correlation coefficients for plaque plus media CSA (0.83 to 0.99 and 0.76 to 0.97, respectively), whereas no influence was observed for external elastic membrane CSA (1.00 to 1.00 and 0.99 to 0.99, respectively). CONCLUSIONS: This study demonstrates the feasibility of this novel algorithm to reduce blood noise, thereby enabling accurate lumen border delineation and providing reproducible measurements of both the lumen and plaque plus media CSAs. Incorporating a digital BNR may serve as an important adjunct to ultra-high-frequency IVUS imaging for improving accurate quantitative evaluation of vessel dimensions.
No takes yet. Share an insight, caveat, or question.
Hibi et al. (2000) conducted an observational in Coronary artery disease (n=35). Blood noise reduction algorithm (BNR) vs. Without BNR (unprocessed images) was evaluated on Intraobserver correlation coefficient for lumen cross-sectional area (CSA). A novel blood noise reduction algorithm improved intraobserver and interobserver correlation coefficients for lumen cross-sectional area in IVUS images.