The influence of emitter current on field desorption (FD) mass spectra was examined with use of an emitter current programmer (ECP). Reproducibility of FD mass spectra measured by electronic detection is improved with ECP and the integration of stored spectra on the computer. The FD mass spectra of some surfactants are shown as an example of complex mixture. The FD mass spectra of poly(oxyethylene) alkylphenyl ethers show that the distribution pattern of molecular ion peaks depends upon emitter current. The fractional desorption also occurs in the FD mass spectra of α-hydro-ω-(octadecylamino)poly(oxyethylene) measured with a constant emitter current. The emitter current required for desorption of hexadecyltrimethylammonium chloride is greatly affected by coexisting with another ammonium salt. Use of ECP supplies information on the analysis of multicomponents.