A high performance liquid chromatography coupled to tandem mass spectrometry (HPLC-MS/MS) method for simultaneous quantification of Δ9-tetrahydrocannabinol (THC), its two metabolites 11-hydroxy-Δ9-tetrahydrocannabinol (11-OH-THC) and 11-nor-9-carboxy-Δ9-tetrahydrocannabinol (THC-COOH), and four additional cannabinoids (cannabidiol (CBD), cannabigerol (CBG), tetrahydrocannabivarin (THCV), and cannabinol (CBN)) in 1 mL of human urine and plasma was developed and validated. The hydrolysis process was studied to ensure complete hydrolysis of glucuronide conjugates and the extraction of a total amount of analytes. Initially, urine and plasma blank samples were spiked with THC-COOH-glucuronide and THC-glucuronide, and four different pretreatment methods were compared: hydrolysis-free method, enzymatic hydrolysis with Escherichia Coli β-glucuronidase, alkaline hydrolysis with 10 M NaOH, and enzyme-alkaline tandem hydrolysis. The last approach assured the maximum efficiencies (close to 100%) for both urine and plasma matrices. Regarding the figures of merit, the limits of detection were below 1 ng/mL for all analytes, the accuracy ranged from 84% to 115%, and both within-day and between-day precision were lower than 12%. Finally, the method was successfully applied to real urine and plasma samples from cannabis users.
No takes yet. Share an insight, caveat, or question.
Aizpurua‐Olaizola et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: