Marginal chordal force averages just 0.006 N in humans, suggesting limited role in preventing mitral prolapse.
Why the study?
The study aimed to quantitatively investigate the role of mitral valve marginal chordae during static closure by analyzing chordal force and its potential correlation with transvalvular pressure.
Marginal chordae experience very low forces during static closure, questioning their traditionally described primary role in preventing mitral prolapse at closure.
Low marginal chordal forces question their primary role in prolapse prevention; hypothesis-generating challenge to consensus needing in vivo validation.
Introduction This study aimed to quantitatively investigate the role of marginal chordae of the mitral valve during static closure by analyzing the force within these chordae and its potential correlation with the transvalvular pressure. Materials and Methods Porcine (n = 10) and human (n = 4) mitral valves were mounted in a saddle-shaped annulus clamp, while an in vitro system was utilized to replicate a physiological systolic blood pressure by vacuum pressure applied to the atrial side of the valve. A Millar pressure catheter and force transducers (“C-gauges”) were used to acquire the data. Strut chordal force was recorded simultaneously with the marginal chordal force, to serve as a verification of the employed equipment. Results The marginal chordal force in porcine and human averaged to 0.012 N ± 0.008 N and 0.006 N ± 0.008 N respectively at peak systolic pressure of 120 mmHg. Additionally, a simplified finite element study was conducted and supports low marginal forces relative to intermediary chords. Discussion & Conclusion This low force calls into question the traditionally-described role of the marginal chordae in preventing mitral prolapse at closure. The chords do, however, still show signs of assisting with the leaflets’ positioning. The relationship between marginal chordal forces and transvalvular pressure varies significantly across experiments, further obfuscating their role in normal valvular function.
No takes yet. Share an insight, caveat, or question.
Ancker et al. (2026) studied Healthy mitral valves (in vitro model) (n=14). Static vacuum pressure (120 mmHg) was evaluated on Marginal chordal force at peak systolic pressure of 120 mmHg. The marginal chordal force in porcine and human mitral valves averaged 0.012 N and 0.006 N respectively at a peak systolic pressure of 120 mmHg, suggesting a limited role in preventing mitral prolapse at closure.