Key result
Finapres arterial pulse wave analysis with Modelflow underestimated cardiac output compared to thermodilution and showed no significant correlation.
Why the study?
Does Finapres arterial pulse wave analysis with Modelflow accurately measure cardiac output compared to invasive thermodilution in healthy elderly subjects?
Observational (n=28)
Does Finapres arterial pulse wave analysis with Modelflow accurately measure cardiac output compared to invasive thermodilution in healthy elderly subjects?
Non-invasive cardiac output measurement using Finapres with Modelflow is inaccurate in healthy elderly subjects and should not be used without invasive calibration.
Answers are grounded in this paper and cite their sources.
May caution against Finapres Modelflow for cardiac output in elderly without calibration; leaves open accuracy in broader populations.
Non-invasive continuous monitoring of cardiac output could be very useful in clinical care and in research settings, particularly in elderly subjects. We studied whether Finapres arterial pulse wave analysis with Modelflow® is a reliable non-invasive method for the assessment of cardiac output in healthy elderly subjects. We compared Modelflow® cardiac output (MFCO) with thermodilution cardiac output (TDCO) in 28 healthy subjects, aged 70±4years (mean±S.D.). TDCO was measured during right-sided heart catheterization, while MFCO was calculated with Modelflow® software from non-invasive arterial Finapres blood pressure, which was measured simultaneously. The two methods were compared using a paired t-test, by Pearson correlation, and by Bland–Altman analysis. TDCO was 6.4±1.1litres/min (mean±S.D.) and MFCO was 4.7±1.3litres/min (P<0.001). There was no significant correlation between MFCO and TDCO (r = 0.28, P = 0.13). Mean difference (bias) was -1.7litres/min (S.E.M. 0.27litres/min), with an S.D. (precision) of 1.5litres/min. The 95% limits of agreement were -4.6 to +1.1litres/min. In conclusion, non-invasive MFCO values differed significantly from and showed no significant correlation with invasively determined TDCO values in the normal range. Although simple, non-invasive and patient-friendly, the Modelflow® method is inaccurate for assessment of cardiac output without invasive calibration.
No takes yet. Share an insight, caveat, or question.
REMMEN et al. (2002) conducted an observational in Healthy elderly subjects (n=28). Finapres arterial pulse wave analysis with Modelflow vs. Thermodilution was evaluated on Cardiac output (litres/min) (MD -1.7, 95% CI -4.6 to +1.1, p=<0.001). Finapres arterial pulse wave analysis with Modelflow underestimated cardiac output compared to thermodilution and showed no significant correlation.