A thin polymeric film, containing a basic polyelectrolyte, was developed with properties that potentially make it highly suitable for use in a membrane blood oxygenator. Its CO2 absorption capacity was correlated as a function of the CO2 partial pressure, temperature and relative humidity of the processed gas. The correlation showed a linear dependence with CO2 partial pressure and relative humidity and an Arrhenius dependence with temperature. A four‐step mechanism for the absorption process was postulated and its parameters were evalutaed. The apparent diffusion constant for CO2 was five times that measured for an ion‐exchange resin made from the polyelectrolyte. The capacity and mechanism correlations obtained are suitable for the design of a CO2 removal unit for a recirculated air atmosphere as well as for a blood oxygenator. The permeability of the film to CO2 and O2 was measured and the ratio was found to be far greater than that of most commercial materials. This suggests that a blood oxygenator utilizing such a film would have the controlling resistance to transfer on the oxygen side, rather than on the blood side, thus perhaps avoiding bubbling on the blood side altogether.