Key result
Pacemaker implantation resulted in a median effective radiation dose of 1.1 mSv, which is substantially lower than doses from coronary angiography (8 mSv) or electrophysiology studies (6 mSv).
Why the study?
What is the patient radiation dose during pacemaker implantations compared to other interventional cardiology procedures?
Observational (n=128)
What is the patient radiation dose during pacemaker implantations compared to other interventional cardiology procedures?
Radiation exposure during pacemaker implantation is relatively low (median effective dose 1.1 mSv) compared to other interventional cardiology procedures.
Pacemaker implantation's low radiation dose may reassure on procedural safety; leaves open prospective outcome comparisons with higher-dose interventions.
The purpose was to evaluate patient radiation doses and compare with other interventional procedures. One hundred and twenty-eight procedures were carried out with a recently installed mobile undercouch C-arm machine with a 23-cm diameter image intensifier. The radiation dose is provided by the X-ray machine, in terms of cumulative dose (CD). Kerma-Area product (KAP) was then estimated from CD and the X-ray field size. Other patient parameters recorded were patient weight, age, kilovolt, milliampere and fluoroscopy time (T). Median (range) CD, KAP and T were 15.2 mGy (3.2-110 mGy), 6.3 Gy cm(2) (1.3-45.7 Gy cm(2)) and 5.2 (1.5-27.4 min) min, respectively. Median E was 1.1 mSv (conversion factor: 0.18 mSv per Gy cm(2)), which corresponds to approximately one lumbar spine X-ray radiography. The effective dose is much lower than a coronary angiography (8 mSv) or an electrophysiology study (6 mSv). Radiation dose is low compared with other interventional cardiology procedures.
No takes yet. Share an insight, caveat, or question.
Tsapaki et al. (2011) conducted an observational in Pacemaker implantation (n=128). Pacemaker implantation vs. Coronary angiography or electrophysiology study was evaluated on Effective radiation dose. Pacemaker implantation resulted in a median effective radiation dose of 1.1 mSv, which is substantially lower than doses from coronary angiography (8 mSv) or electrophysiology studies (6 mSv).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: