Key result
The new AlluraClarity X-ray technology reduced radiation exposure by 69% for non-CRT and 75% for CRT pacemaker and ICD implantations compared to reference technology.
Why the study?
Does new X-ray technology reduce radiation exposure in patients undergoing pacemaker and ICD implantation?
Cohort (n=1,185)
No
Does new X-ray technology reduce radiation exposure in patients undergoing pacemaker and ICD implantation?
Answers are grounded in this paper and cite their sources.
The use of new X-ray technology with advanced image processing algorithms significantly reduces patient radiation exposure during pacemaker and ICD implantations by 69-75% without compromising procedure time or contrast volume.
New X-ray technology providing new image processing techniques may reduce radiation exposure. The aim of this study was to quantify this radiation exposure reduction for patients during pacemaker and implantable cardioverter defibrillator (ICD) implantation. In this retrospective study, 1185 consecutive patients who had undergone de novo pacemaker or ICD implantation during a 2-year period were included. All implantations in the first year were performed using the reference technology (Allura Xper), whereas in the second year, the new X-ray technology (AlluraClarity) was used. Radiation exposure, expressed as the dose area product (DAP), was compared between the two time periods to determine the radiation exposure reduction for pacemaker and ICD implantations without cardiac resynchronization therapy (CRT) and with CRT. Procedure duration and contrast volume were used as measures to compare complexity and image quality. The study population consisted of 591 patients who had undergone an implantation using the reference technology, and 594 patients with the new X-ray technology. The two groups did not differ in age, gender, or body mass index. The DAP decreased with 69 % from 16.4 ± 18.5 to 5.2 ± 6.6 Gy cm 2 for the non-CRT implantations ( p < 0.001). The DAP decreased with 75 % from 72.1 ± 60.0 to 17.8 ± 17.4 Gy cm 2 for the CRT implantations ( p < 0.001). Nevertheless, procedure duration and contrast volume did not differ when using the new technology ( p = 0.09 and p = 0.20, respectively). Introduction of new X-ray technology resulted in a radiation exposure reduction of more than 69 % for patients during pacemaker and ICD implantation while image quality was unaffected.
No takes yet. Share an insight, caveat, or question.
Dijk et al. (2016) conducted a cohort in Pacemaker or ICD implantation (n=1,185). New X-ray technology (AlluraClarity) vs. Reference X-ray technology (Allura Xper) was evaluated on Radiation exposure (dose area product) for non-CRT implantations (69% reduction, p=<0.001). The new AlluraClarity X-ray technology reduced radiation exposure by 69% for non-CRT and 75% for CRT pacemaker and ICD implantations compared to reference technology.