We have constructed a diamond anvil cell and a cryogenic system for high-pressure experiments at temperatures down to 1.5 K. A new mechanism is used to apply the force, and the body of the cell contains a heat-exchanger for use with liquid helium or liquid nitrogen. With the press, a maximum force of 30 000 N can be applied to the diamonds. The force can be changed at any temperature from outside the cryostat. We describe a technique to load metal samples with hydrogen at high pressures and do in situ resistance measurements. With this technique, we have synthesized stoichiometric PdH and PdD at a pressure of ∼4 GPa. We also have measured the pressure dependence of their superconducting transition temperatures. For PdH we find Tc=8.8 K and ∂ ln Tc/∂P=−6.8×10−2 GPa−1, and for PdD we find Tc=11.05 K and ∂ ln Tc/∂P=−5.0×10−2 GPa−1.
No takes yet. Share an insight, caveat, or question.
Hemmes et al. (1989) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:
More from Review of Scientific Instruments