The wavelength shift with pressure of the ruby R1 fluorescence line (Δλ) has been calibrated in the diamond-window pressure cell from 0.06 to 1 Mbar. This was done by simultaneously making specific volume measurements of four metals (Cu, Mo, Ag, and Pd) and referring these results to isothermal equations of state derived from shock-wave experiments. The result is P (Mbar) = (19.04/5) {[(λ0+Δλ)/λ0]5−1}, where λ0 is the wavelength measured at 1 bar.
No takes yet. Share an insight, caveat, or question.
Mao et al. (1978) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: