Key result
The A&D BP UB-543 wrist device met the validation requirements of the European Society of Hypertension International Protocol, overestimating systolic BP by 1.1±2.9 mmHg and underestimating diastolic BP by 0.1±3.0 mmHg.
Why the study?
Is the A&D BP UB-543 wrist device accurate for home blood pressure measurement according to the European Society of Hypertension International Protocol?
Population
33 adult patients
Comparison
A&D BP UB-543 wrist device for home blood… vs Reference standard according to the European…
Design
Other
Authors
Loading...
Validates A&D UB-543 wrist device accuracy for home BP monitoring; leaves open need for outcome trials before routine clinical adoption.
Observational (n=33)
Is the A&D BP UB-543 wrist device accurate for home blood pressure measurement according to the European Society of Hypertension International Protocol?
The A&D BP UB-543 wrist device with correct position guidance meets the ESH International Protocol requirements and is accurate for clinical use in adults.
Fania et al. (2015) conducted an observational in Blood pressure measurement (n=33). A&D BP UB-543 wrist device vs. Standard protocol requirements was evaluated on Accuracy of systolic and diastolic BP measurement. The A&D BP UB-543 wrist device met the validation requirements of the European Society of Hypertension International Protocol, overestimating systolic BP by 1.1±2.9 mmHg and underestimating diastolic BP by 0.1±3.0 mmHg.