Key result
The A&D BP UB-542 wrist device passed all ESH protocol requirements, overestimating systolic BP by an average of 1.8±7.2 mmHg and diastolic BP by 1.6±5.7 mmHg.
Why the study?
Is the A&D BP UB-542 wrist device accurate for home blood pressure measurement according to the ESH International Protocol?
Population
33 adult patients, mean age 50.9 years, mean systolic BP 141.6 mmHg, mean diastolic BP 89.2 mmHg.
Design
Other
Authors
Loading...
May support home BP monitoring per ESH criteria; leaves open need for outcome studies in broader populations.
Observational (n=33)
No
Is the A&D BP UB-542 wrist device accurate for home blood pressure measurement according to the ESH International Protocol?
Mean Difference: 1.8
The A&D BP UB-542 wrist device meets the ESH International Protocol requirements and is valid for clinical use in adults.
Saladini et al. (2013) conducted an observational in Blood pressure measurement (n=33). A&D BP UB-542 wrist device vs. Standard measurement (ESH protocol) was evaluated on Accuracy of BP measurement according to ESH International Protocol (Mean difference 1.8±7.2 mmHg (systolic) and 1.6±5.7 mmHg (diastolic)). The A&D BP UB-542 wrist device passed all ESH protocol requirements, overestimating systolic BP by an average of 1.8±7.2 mmHg and diastolic BP by 1.6±5.7 mmHg.