Introduction The accuracy of restorations fabricated through digital workflows is partially determined by the intraoral scanner utilized. This study evaluated the marginal fit of digitally fabricated lithium disilicate single crowns produced using two intraoral scanners: the iTero® Element and the Planmeca® Emerald. Methods Eleven non-carious molar teeth were prepared for full-coverage lithium disilicate crowns and scanned with both intraoral scanners. Restorations were designed using Romexis software and milled with a Planmeca® milling unit. Marginal fit, defined as the vertical discrepancy between the restoration margin and the prepared tooth surface, was measured at six locations per crown using a digital microscope at 80× magnification. Results and discussion Crowns fabricated from iTero® scans showed more consistent measurements, whereas those from Planmeca® scans showed greater variability. The mean marginal discrepancy was 51.27 µm for the iTero® group and 73.18 µm for the Planmeca® group. Although the iTero® group exhibited a lower mean marginal discrepancy, the difference between groups was statistically significant ( p = 0.032). All restorations were within the clinically acceptable marginal discrepancy range (<120 µm). Within the limitations of this in vitro study, both intraoral scanners produced lithium disilicate crowns with clinically acceptable marginal fit, and statistically significant difference was observed between the two systems.