Key result
Doppler echocardiography successfully determined the optimal atrioventricular interval for DDD pacing, showing a significant linear relationship with the predicted optimal interval (r=0.70, P<0.01).
Why the study?
Can Doppler echocardiography determine the optimal atrioventricular interval in patients with complete atrioventricular block and DDD pacemakers?
Can Doppler echocardiography determine the optimal atrioventricular interval in patients with complete atrioventricular block and DDD pacemakers?
Doppler echocardiography provides a noninvasive, repeatable, and simple approach to successfully determine the optimal atrioventricular interval setting for DDD pacing.
BACKGROUND: We investigated the clinical value of Doppler echocardiography for determining the optimal setting of atrioventricular interval in patients with dual chamber pacemakers (DDD). In 38 patients with complete atrioventricular block and DDD pacemakers, the atrioventricular interval was prolonged in a stepwise fashion by 20 ms, from 90 to 250 ms, and cardiac stroke volume and transmitral flow were measured by pulsed Doppler echocardiography. METHODS: The optimal atrioventricular interval at which the cardiac stroke volume was maximal was 168.9 +/- 15.6 ms. The atrioventricular interval was 178.4 +/- 23.4 ms when the end of the A wave coincided with complete closure of the mitral valve. RESULTS: There was a significant linear relationship between the optimal atrioventricular interval and the predicted optimal atrioventricular interval (Y = 86.2 +/- 0.5X, r = 0.70, standard error of the estimate [SEE] = 11.5, P < 0.01). CONCLUSION: The optimal atrioventricular interval setting for DDD pacing can be successfully determined using Doppler echocardiography, which is a noninvasive, repeatable, and simple approach.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2009) studied Complete atrioventricular block (n=38). Doppler echocardiography for determining optimal atrioventricular interval was evaluated on Relationship between optimal atrioventricular interval and predicted optimal atrioventricular interval (r = 0.70, p=< 0.01). Doppler echocardiography successfully determined the optimal atrioventricular interval for DDD pacing, showing a significant linear relationship with the predicted optimal interval (r=0.70, P<0.01).